서브 비주얼

Thank you for visiting ATS

ADVANCED TECHNOLOGY SOLUTION

컨텐츠

Air Leak Tester

I-28 Leak Tester

LIST

Product Features
Standard Test Types :
▶ Pressure/vacuum decay : leak standard
▶ Pressure/vacuum decay : △P
▶ Pressure/vacuum decay : △P/△T
▶ Pressure/vacuum occlusion
▶ Pressure mass flow
▶ Pressure mass flow - leak standard
▶ Parent program linking
Product Information

99 Test Programs

Application Flexibility including individual user programmed:

 Test types

 Timers

 Pressure limits

 Reject limits

 Calibration parameters

 Units of measurement

 Digital I/O

 Tooling control  

 

Patented Auto Setup

 Automated optimization of test program based on maximum user allowable cycle time

• Significantly simplifies instrument test programming and setup

 

32-bit Processor & 24-Bit A/D Converter

• Exceptionally fast, high resolution test processing 

• Stable yet extremely responsive pressure/flow measurements

 

Absolute Pressure Transducer

• Eliminates test variations from barometric pressure changes 

• Uncertainty 0.05% full scale (all ranges)

 

Pressure Range

• Standard from vacuum to 500.0 psig 

• Option - above 500 psig

 

Standard Mass Flow Ranges

• Standard ranges from 0.5 to 3,000 sccm 

• Above 3,000 sccm available

• Uncertainty 0.5% full scale (standard ranges)

 

Transducer Verification/Recertification

• Easy pressure or flow transducers 

• 6-point standard pressure calibration

• 2-point standard flow calibration

 

Available Pneumatic Manifolds-Optimization

• Standard flow 

• Low volume

• High flow

 

Test Ports

• Choice of: 1/4" FNPT or 3/8" FNPT

• Other sizes available

 

Pressure Control

• Standard high precision manual regulators 

• Two standard regulators for mass flow

• Select up to two precision electronic regulators for automatic pressure changes unique to test part or test sequence

• Six point regulator Calibration

• Optional internal vacuum generator for vacuum applications

 

Units of Measure

• All standard industrial units of measure are easily selectable per part program

 

Self-Test Functions

• Internal leak detection process

• Program calibration verification

 

Leak Standards

• Internal Leak Standard on the valve manifold for pressure decay with orifice configurations

 

RS232 Communication Ports

• 2-Way Telnet communication

• Test result data transmission with definable fields

• Pressure streaming for curve storage

• Email of reports, test data ,and alerts

• Barcode unique part identification

 

Ethernet Port

• 2-Way Telnet communication w/Email of reports, test data ,and alerts

• EtherNet/IP™ Option 26 Inputs/25 Outputs

• Test result data transmission with definable fields

• Test Program Selection

 

USB Port

• Backup/Restore

• Cloning

• Report storage

• Test result data storage & result synchronization

 

Automatic Calibration

• Easy to perform

• Permits manual edits if averaging multiple parts is required

• Calibrates to master part

 

Environmental Drift Correction

• Automatically maintains calibration accuracy to compensate for changes in environmental conditions including temperature

 

Quik Test Function

• Shortens test cycle when it is obvious and imminent the monitored pressure curve will result in a reject or accept

• Analyzes test results in real time

• Provides projected final test results

 

24 Volt Digital Inputs/Outputs

• 12 User configurable inputs

• 12 User configurable outputs

• Tooling control – up to five tooling motions with feedback - eliminates the need for a PLC in some cases

 

Key or Password Security

• Select menu items to secure or unsecure

 

Standard Selectable Menu Languages

• English

• Spanish

• Chinese

 Korean

• Portuguese

 

Data management & Storage

• More than 5,000 tests stored on internal memory

• Infinite - Expandable Storage Via USB

• Statistic Data Tracking

• Tracks Percent of Accept and Rejects

• Provides Average Accept and Reject

• Standard Deviation on Accepts

• Allows definable Sample Size

• Resettable counters:

• Test Results Including Malfunction

• Data Viewable on Display

 

3-Color Light-Bar Indicator

• Unique panel mounted light indicators provides clear feedback of each test

• Eliminates stack lights

• White: In Test

• Green: Accept

• Red: Reject

• Adjustable brightness

• Adjustable lighting duration

 

Compact Enclosure Design

• Dimensions: 9.25" high x 12" wide x 8.75" deep