Calibrated Leak Standards
Flow Range
• Leak or Flow range: 1x10-5 atm-cc/sec to 50,000 scc/min - Manufactured to ISO 17025
Pressure Range
• Standard holders: -14.7 to 500 psig (-1.0 to 34.5 bar)
• Welded holders: >500 to 3,000 psig (>34.5 to 206.8 bar)
Accuracy and Manufacturing tolerance
• Standard Flow Range: 10.0 scc/min or greater
- Flow Accuracy ±1.4%
- Manufacturing Tolerance ±1.0%
• Low Flow Range: 0.5 to 10.0 scc/min
- Flow Accuracy ±1.4% or ±0.1 scc/min (whichever is greater)
- Manufacturing Tolerance ±0.1 scc/min
• Special Flow Range: 0.49 scc/min or Less
- Flow Accuracy ±6.5%
- Manufacturing Tolerance ±20%
Integral Assembly
The manufactured flow standard is sealed into one of various assemblies:
• General application (1/8" NPT end fittings)
• Legacy units for CTS
• New units for and LPC 850, CTS Sentinel I28, C28, C20WE, LPC 528, 3520, Blackbelt, and Blackbelt Pro instruments
• Compatible replacement units for competitor instruments
• Custom mounting configurations
All leak standard assemblies include an inlet filter, some models are equipped with an exhaust filter.
Integral Assembly
• For re-certification, the current flow rate is measured and compared to previous calibration and/or the original leak rate and pressure specification. If the current flow value is outside the allowable toleranace range it is replaced with a new orifice. All CTS recalibrated leak standards come with a NIST traceable certificate and specifications.